Customization: | Available |
---|---|
Certification: | CE, ISO |
Place Style: | Vertical |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
High frequency induction heating furnace![]() |
working voltage: 220VAC, single phase working current: 2-52a (transformer transformer current) (requires 63A air switch) output power: 25KW output frequency: 30-80khz (adjustable) cycle ratio: 100% copper coil size: 170mm (OD) x 160mm (ID) x 90mm (H) protection system: automatic water pressure, overheat, overpower protection refractory materials: equipped with multi-layer graphite fiber layer and alumina fiber layer to support graphite crucible and prevent thermal radiation |
Temperature control system and thermocouple![]() |
Working voltage: 208-240vac, single phase. PID control can be used to manually adjust the current output. The equipment is overheat and broken. · Temperature control accuracy: + / -5 (thermocouple probe located at the top of the sample) The highest working temperature: 2200 ºC (working time is less than 30 min) Long working temperature: 200 pieces C ~ 2100 Maximum heating rate: The equipment is overheat and broken Temperature control accuracy: + / -5 (thermocouple probe located at the top of the sample) The highest working temperature: 2200 ºC (working time is less than 30 min) Long working temperature: 200 pieces C ~ 2100 Maximum heating rate: Temperature control accuracy: + / -5 (thermocouple probe located at the top of the sample) The highest working temperature: 2200 ºC (working time is less than 30 min) Long working temperature: 200 pieces C ~ 2100 Maximum heating rate: 10ºC/second (200ºC-1200ºC) 8ºC/second (1200ºC~1500ºC) 2ºC~4ºC/second (1500ºC~2200ºC) |
Graphite crucible and firebrick![]() |
The instrument is equipped with a high-purity graphite crucible: 58mm(OD)×48mm(ID)×94mm Maximum capacity: the sample size shall be 2/3 of the volume of graphite crucible Instruments equipped with resistance to 2500 ºC in the high purity graphite fiber felt and the alumina fiber felt can be resistant to 1700 degrees, Used for lifting graphite and preventing thermal radiation effect flange sealing |
Seal flange and quartz tube |
(when the temperature is above 1700 degrees) flange adopts water cooled flange and is installed with quartz observation port for the light of the infrared thermometer.The flange is fitted with a stainless steel needle valve.(infrared thermometer) Bottom flange: install stainless steel needle valve and KF25 interface (connected with vacuum pump) quartz tube size 150mm (OD) x142mm (ID) x 300mm (L) |
vacuum gauge![]() |
The entire vacuum degree is free of deviation, accuracy and sensitivity. |
Circulating water cooler ![]() |
Temperature range: 5-30 ºC Cooling water flow rate: 16L/min Water tank: 12L, made of stainless steel Working voltage: ac208-240v, single phase Model: CW5300 |
Vacuum pump![]() |
The instrument is equipped with a double - rotating mechanical pump vrd-8 Extraction rate: 8 m3 / h, limit vacuum degree 6X10 e-2 Pa Working voltage: ac208-240v, single phase Maximum power: 375W In order to obtain a higher vacuum degree to achieve better smelting effect, the vortex molecular pump can be selected and the vacuum degree can reach 10e-4pa |
product size | induction melting furnace: Main Volume (CM):55.5L×25.2L×46.2H extension set (CM):36×26×36 Circulating water cooler:460mm(L) x 475mm(W) x 762mm(H) Moving frame size:600mm(L) x 1000mm(W) x 600mm(H) |