Model | KBF13Q-I | KBF13Q-II | KBF13Q-III | KBF13Q-IV | KBF13Q-V |
Power |
3 KW | 6 KW | 8 KW | 10 KW | 13 KW |
Inside Chamber Size |
Heated chamber: 160x150x150mm, 3.6 Liter. |
Heated chamber: 200x200x200mm, 8 Liter. |
Heated chamber: 300x200x200mm, 12 Liter. |
Heated chamber: 300x250x250, 19 Liter. |
Heated chamber: 400x300x300mm, 36 Liter. |
Product Dimensions (mm) | 550x470x860 | 600x510x910 | 600x610x940 | 650x610x990 | 700x710x1355 |
Furnace Structure | 1. Double layer stainless steel structure for maximum vacuum and safety, which can ensure a positive pressure up to 0.06MP and vacuum up to -0.09Mpa 2. Spring loaded pressure relief valve installed on the furnace provide protection against chamber over-pressurization. 3. Water cooling front door and air cooling shell. 4. Built in the vacuum pump and two-channel gas flowmeters, pressure gauge needle valves. |
||||
Inside Chamber Size | 1. Energy saving 1800 grade fibrous alumina insulation. 2. High purity Al2O3 coating. |
||||
Voltage | AC 220 - 380V single phase; 50/60 Hz | ||||
Continuous Working Temperature | 1650°C | ||||
Maximum Working Temperature | 1700°C (< 1 hr) | ||||
Heating Rate | 20°C / min | ||||
Temperature Accuracy | +/- 1°C | ||||
Heating Elements | 1, 6 pcs Mo Alloy coils (can only be used under inert gas, Ar is preferred; cannot be used under air or oxygen at the temperature > 300°C). 2,Optional: Heating elements can be replaced by MoSi2 to enable heat processing under oxygen and air atmospheres up to 1700°C. |
||||
Temperature Control | 1. PID automatic control via SCR (Silicon Controlled Rectifier) power control with phase angle fired, current limiting resistor. 2. a40 programmable segments for precise control of heating / cooling rate and dwell time. 3. Over temperature protection and thermocouple failure alarm functions allow operation without attendant(s). 4. +/- 1ºC temperature accuracy. 5. RS485 Communications Port. |
||||
Pressure Relief Valve | Spring Loaded Type Pressure Relief Valve • Cracking pressure is manufacturer pre-set at 0.066Mpa (relative pressure value). • Valve close pressure: 0.03Mpa (relative pressure value). • Max. Temperature 200ºC. • 35mm O.D barb fitting for gas exhaust. |
||||
Gas Flow Control and Vacuum Valve |
• Two flow meters (#1: 0~8L/min; #2: 0~4L/min) are installed on the front panel of the furnace. • One heavy duty vacuum pump (CE certified) is built on the bottom of the furnace. |
||||
Water Cooling | • A 16L/min recirculating water chiller is installed in bottom of furnace for cooling furnace sealing flange. | ||||
Warranty | One year limited warranty with lifetime support. | ||||
Application Notes | • For preserving the longevity of the furnace components, we recommend the heating rate not to exceed 5 °C /min. • The cooling rate shall also not exceed 5 °C /min. • Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating. • The refractory block must be inserted before closing door. |