Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Power Supply: | 110/220V |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Working Voltage | 220V AC, 50/60Hz, Single Phase |
Max. Power Consumption | 3KW |
Glove Box Chamber | · Case Material: Stainless Steel 304 · Dimensions: 2400mm(L) x 800mm(W) x 930mm(H) · 12mm thick double layer laminated safety glass with corrosion protection film for operating in a safe and easy manner |
Chamber Condition |
· Water concentration: <1 ppm (20°C, 1 atm) · Oxygen concentration: <1 ppm (20°C, 1 atm) |
Leak rate | <0.05 Vol% / hr (Depending on oxygen concentration inside the glove box) |
Air-lock Chamber |
· Antechamber Dimensions: 385mm(Dia.) x 588mm(L) · Small Chamber Dimensions (for tools): 150mm(Dia.) x 300mm(L) · Max. vacuum level: 100 Pa |
Working Gas |
· Operating gas: Inert gas such as N2, Ar, He · Regenerating gas: Mixture of H2 (10%) and Operating gas o For purification system with only water adsorbing function, Regenerating gas will be the same as Operating gas |
Gas Purification System |
· Automated Regeneration Process · Automated water and oxygen removal · Purification Cylinder: made of BASF antioxidant materials and UOP water adsorbent materials · Moisture removal capacity: 1.5KG · Oxygen removal capacity: 180L |
Filter System |
· Organic Solvent Filter System filled with replaceable activated carbon · Built in HEPA filter can eliminate particle size > 0.3μm |
Control system | · Siemens PLC control system with 6" touch screen display o Water concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm o Oxygen concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm o Pressure: -2500 to 2500 Pa with an accuracy of 1 Pa |
Pump |
· England EDWARDS pump is included for faster vacuum and removing moisture · Vacuum rate: 8.4 - 12 m^3/hr |
Blower | Flow Rate: 90m³/h |
Purification Pipeline | 304 Stainless Steel, KF40 Port |
Gloves | A pair of NORTH butyl glove is included to provide better protection against toxic materials |
Lighting | Built in fluorescent lamp |
Slidable Shelf |
· Three slidable shelves are built 0n top of inside glove chamber with dimensions of 300mm x 250mm · 300mm away from roof |
Product Dimensions | 3115mm(L) x 830mm(W) x 1830mm(H) (123" x 33" x 72") |
Shipping Dimensions & Weight | Total 2 pallets #1: 86"x41"x84" #2: 86"x41"x83" Total weight 1800 lb |
Warranty |
· One Year limited warranty with lifetime support · Rusting and damage due to improper storage condition or maintenance is not covered by warranty · Gloves are consumable items and are not covered under warranty |
Application Notes | · The channel between the glove box chamber and the gas purification system must be unimpeded during the operation of gas purifying · Always check the joints, gloves, sealing elements, and pipelines for air-tightness before and during the operation · Corrosive gases are prohibited as they will damage the water sensor and oxygen sensor · The inlet and outlet of the Air-lock Chamber must not be opened at the same time to prevent contamination by the external gas · Coolant water outlet must remain switched on all the time · System check frequently to replace filters and solvent adsorbents for maintanence and maximum efficiency |
Specifications of Standard Model of Purification Glove Box | ||||||
No. | Model No. | Box Dimensions | Total Dimensions | Operating Sides | Numbers of Gloves | Options |
1 | GBP800S-2 | 1200*800*930mm | 1915*830*1830mm | Single | 2 | Single or double purification columns |
2 | GBP1000S-4 | 1200*1000*930mm | 1915*1060*1830mm | Single | 4 | Single or double purification columns |
3 | GBP1000D-4 | 1200*1000*930mm | 1915*1060*1830mm | Double | 4 | Single or double purification columns |
4 | GBP1200S-2 | 1200*1200*930mm | 1915*1260*1830mm | Single | 2 | Single or double purification columns |
5 | GBP1200D-4 | 1200*1200*930mm | 1915*1260*1830mm | Double | 4 | Single or double purification columns |
6 | GBP800S-3 | 1500*800*930mm | 2215*1060*1830mm | Single | 3 | Single or double purification columns |
7 | GBP1000S-3 | 1500*1000*930mm | 2215*1060*1830mm | Single | 3 | Single or double purification columns |
8 | GBP1000D-6 | 1500*1000*930mm | 2215*1060*1830mm | Double | 6 | Single or double purification columns |
9 | GBP1200S-3 | 1500*1200*930mm | 2215*1260*1830mm | Single | 3 | Single or double purification columns |
10 | GBP1200D-6 | 1500*1200*930mm | 2215*1260*1830mm | Double | 6 | Single or double purification columns |
11 | GBP800S-4 | 1900*800*930mm | 2615*1060*1830mm | Single | 4 | Single or double purification columns |
12 | GBP1000S-4 | 1900*1000*930mm | 2615*1060*1830mm | Single | 4 | Single or double purification columns |
13 | GBP1000D-8 | 1900*1000*930mm | 2615*1060*1830mm | Double | 8 | Single or double purification columns |
14 | GBP1200S-4 | 1900*1200*930mm | 2615*1260*1830mm | Single | 4 | Single or double purification columns |
15 | GBP1200D-8 | 1900*1200*930mm | 2615*1260*1830mm | Double | 8 | Single or double purification columns |
Notes: Cabinet water and oxygen content under the standard condition: Water content ≤1ppm, oxygen content≤1 ppm (Note: Standard condition is that an atmospheric pressure, constant temperature of 20 degree). Leakage Rate: ≤0.05Vol%/ h (subject to change value of oxygen content in box) Used Gases a) Working Gas: N2,Ar,He and etc. b) Controlling Gas: Compressed air or Inert Gas. C) Restoring Gas: Mixed gas of working gas and hydrogen. Restoring gas is as same as working gas if there is only water removal system or oxygen removal system. Remarks: All Single side boxes of 4 gloves are assembled glove box. Numbers of assembled box can be ordered as per customer's needs. Quantity of purification columns is also designed to be matched according to customer's requirements |