Customization: | Available |
---|---|
Application: | Industry, School, Hospital, Lab |
Customized: | Customized |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model | 1100CAtmosphere-I | 1100CAtmosphere-II | 1100CAtmosphere-III |
Power | 4KW | 9KW | 16KW |
Furnace structure | 1. The double shell structure is adopted and the fan is cooled. 2. Imported 310S stainless steel materials are imported. 3. Install the water cooling system at the door of the furnace and reduce the sealing temperature of the o-ring. 4. Installation of combustion system (including: igniter, combustion controller, burner). 5. Installation of explosion-proof valve (free setting valve opening pressure). 6. The installation of hydrogen detector can detect the leakage of hydrogen. |
||
Chamber size(mm) | 300X200X200 | 380X280X280 | 480X380X380 |
Overall size(mm) | 480X590X 712 | 580X690X930 | 680X 790X950 |
Voltage | 220V | 380V | |
Phase | Single | three | |
Heating element | Swedish Canthal A1 import resistance wire, the surface temperature up to 1420 ºC | ||
Control system | The equipment is labeled with Japan's imported FP93 temperature control instrument . Adopt the yu electric program control thermometer 1. 30 program control temperature intelligent PID adjustment. 2, over temperature protection is broken accidentally protection with a temperature or broken accidentally electric furnace heating circuit automatically cut OFF, and, when the temperature of the furnace more than 1200 degrees or thermocouple burn out, the main circuit of ac relay will automatically disconnect, main circuit disconnect, panel ON lights, OFF the light, the limited protection electric furnace). 3.With a 485 communication interface (the selection of the software for the selection of software) 4. It has the function of power failure protection, that is, when the power is off, the program will not start up from the beginning temperature, but the furnace will start to rise when the power is off. 5. The instrument has the function of temperature self-tuning Imported instrument (optional) FP93SHIMADEN, Japan EUROTHERM Touchscreen |
||
Chamber | 1. High quality high pure alumina polycrystalline fiber cured furnace with vacuum suction and filter molding. 2. Adopt Japanese technology to form. The spacing and pitch of the resistance wire in the furnace are arranged according to the best thermal technology of Japan, and the temperature field is simulated by the hot working software 4. Using 4 sides heating (left and right sides), the temperature field is more balanced |
||
Control precision | +/- 10 ºC | ||
Trigger | Phase-shifted trigger | ||
electric appliance | Zhejiang Zhengtai | ||
silicon controlled | 106/16E SEMIKRON, Germany | ||
maximum temperature | 1200 ºC | ||
Rated temperature (working long hours) | 1100 ºC | ||
heating rate | ≤20 ºC/ Min (adjustable) | ||
Recommended heating rate | ≤10 ºC/ Min | ||
· Thermocouple Type | K | ||
vacuum degree | In cold state, it can reach -0.1 MPa | ||
· position limit switch | The power will turn off automatically | ||
Internal materiel | 4mm nickel base alloy |